NMR Chemical Shift Converter

Chemistry

Spectrometer

MHz

Linked to the ¹H frequency through γ/2π = 42.577 MHz/T.

TMS / DSS (organic / aqueous; one ¹H scale)

MHz

The 0-ppm frequency of the observed nucleus from your acquisition parameters (on Bruker, SF).

¹H on a 400 MHz (≈ 9.395 T) magnet: 0 ppm = 400 MHz · 1 ppm = 400.000 Hz

Separate values with new lines, commas, semicolons, spaces or tabs. Commas are always separators, never decimal marks.

Results (¹H, 1 ppm = 400.000 Hz)

#δ (ppm)Δν (Hz)ν (MHz)τ (historical)
14.13841655.36400.00165545.8616
24.12061648.24400.00164825.8794
34.10281641.12400.00164115.8972
44.08501634.00400.00163405.9150

Spacings

4.1384 → 4.1206: 0.0178 ppm = 7.12 Hz

4.1206 → 4.1028: 0.0178 ppm = 7.12 Hz

4.1028 → 4.0850: 0.0178 ppm = 7.12 Hz

Mean gap 7.12 Hz · centre 4.1117 ppm

In a first-order multiplet, equal spacings are the coupling constant J.

Dual-scale ruler

¹H overview: 12 → 0 ppmzoom4.094.104.114.124.134.14ppm16351640164516501655Hz7.12 Hz7.12 Hz7.12 Hz4.13844.12064.10284.0850

Hover over or focus a stick to read its position.

Upper axis: δ in ppm, decreasing to the right. Lower axis: Hz from 0 ppm at 400 MHz. Brackets give the gaps between neighbours in Hz.

About This Tool

NMR Chemical Shift Converter – ppm, Hz and MHz for Any Nucleus and Field

Every NMR spectrum mixes two kinds of numbers. Chemical shifts are reported in parts per million, while coupling constants, linewidths and the spectrometer itself are described in hertz and megahertz. This NMR chemical shift converter moves between the two for ¹H, ¹³C, ¹⁵N, ¹⁹F, ³¹P and a dozen other nuclei at any field strength, measures J from a peak list, predicts how an AB pattern changes with the magnet, and re-references spectra to the solvent signal.

Why shifts are in ppm and couplings in Hz

The shielding that produces a chemical shift is proportional to the applied field, so a peak moves twice as many hertz on a magnet twice as strong. Dividing by the spectrometer frequency removes that dependence and gives a number, δ, that is the same on every instrument. A J coupling is transmitted through the bonding electrons and does not depend on the field at all. That is why a doublet that spans 0.13 ppm at 60 MHz spans only 0.013 ppm at 600 MHz, while J stays at 8 Hz.

The δ definition and a worked ppm to Hz example

IUPAC defines δ = (ν − ν_ref) / ν_ref × 10⁶, where ν_ref is the frequency of the 0-ppm reference for the observed nucleus in the same magnet. Once that frequency, SF, is known the conversion is exact: Δν (Hz) = δ (ppm) × SF (MHz). On a 400 MHz spectrometer the CHCl₃ signal at 7.26 ppm lies 7.26 × 400 = 2904 Hz above TMS. The CDCl₃ carbon at 77.16 ppm on the same magnet is 77.16 × 100.58 = 7760.8 Hz from TMS, because ¹³C is observed at 100.58 MHz, not at 400 MHz.

1 ppm equals the observe frequency in Hz
So it depends on the nucleus, not on the instrument's ¹H rating. On a "400 MHz" magnet 1 ppm is 400 Hz for ¹H, 100.58 Hz for ¹³C, 376.38 Hz for ¹⁹F and 40.55 Hz for ¹⁵N.

The IUPAC unified scale Ξ

The 2001 IUPAC recommendations tabulate, for each nucleus, the ratio Ξ of its 0-ppm frequency to the ¹H frequency of TMS in the same field. The calculator uses SF_X = SF_H × Ξ / 100 for every nucleus, so one table defines where 0 ppm falls on any magnet: 25.145020 % for ¹³C, 40.480742 % for ³¹P and 94.094011 % for ¹⁹F. If you copy the exact SF from your acquisition parameters, the tool uses it instead.

Why higher fields simplify spectra

Two coupled protons form an AB system. When their shift difference in Hz, Δν, is large compared with J, you see two clean doublets (AX). When Δν/J is small, the inner lines grow, the outer lines fade and the doublets lean towards each other: the roof effect. Because Δν scales with the field and J does not, the same molecule that gives a tangled second-order pattern at 60 MHz can look first-order at 600 MHz. The stacked spectra in the coupling tab show this directly, and the analysis mode recovers J and the true shifts from four measured AB lines.

Referencing to residual solvent signals

Most organic spectra are calibrated on the residual protio signal of the deuterated solvent. Accepted values include 7.26 ppm for CHCl₃ in CDCl₃, 2.50 ppm for DMSO-d₆ and 77.16 ppm for the CDCl₃ carbon. If the solvent peak appears at the wrong position, the same offset is added to every peak. The HDO signal of D₂O moves with temperature, so DSS is the better reference for aqueous samples.

The ¹⁵N reference split

Chemists reference ¹⁵N to nitromethane, the IUPAC primary standard, while protein NMR uses liquid ammonia, about 380 ppm lower. The change of reference is a ratio of frequencies, not a fixed offset: the familiar 380.44 ppm is exact only at δ = 0 and drifts by about 0.1 ppm across the amide region. The re-reference tab shows the exact value next to the constant-offset approximation.

The historical τ scale

Older literature uses the τ scale introduced by Tiers in 1958, τ = 10 − δ, on which TMS sits at 10 and shifts increase upfield. It is shown as an extra column for ¹H so old values can be compared with modern δ values.

Frequently Asked Questions

Is the NMR Chemical Shift Converter free?

Yes, NMR Chemical Shift Converter is totally free :)

Can I use the NMR Chemical Shift Converter offline?

Yes, you can install the webapp as PWA.

Is it safe to use NMR Chemical Shift Converter?

Yes, any data related to NMR Chemical Shift Converter only stored in your browser (if storage required). You can simply clear browser cache to clear all the stored data. We do not store any data on server.

How do I convert ppm to Hz in NMR?

Multiply the shift in ppm by the spectrometer frequency of the observed nucleus in MHz: Δν (Hz) = δ (ppm) × SF (MHz). On a 400 MHz instrument a ¹H signal at 7.26 ppm lies 7.26 × 400 = 2904 Hz from TMS. To go back, divide the Hz value by the same frequency.

Why is 1 ppm not 400 Hz in a ¹³C spectrum from a 400 MHz spectrometer?

The 400 MHz rating is the ¹H frequency. Carbon-13 resonates at about 25.145 % of it, so the same magnet observes ¹³C at 100.58 MHz and 1 ppm of ¹³C is 100.58 Hz. Always multiply by the observed nucleus's own frequency, never by the ¹H rating.

Why are coupling constants given in Hz but chemical shifts in ppm?

A chemical shift grows in proportion to the magnetic field, so dividing by the spectrometer frequency (ppm) makes it the same on every instrument. A coupling constant J comes from the bonding electrons and does not depend on the field at all, so it is quoted in Hz. That is why multiplets look narrower in ppm on a stronger magnet.

What magnetic field strength is a 600 MHz NMR spectrometer?

Divide the ¹H frequency by the proton gyromagnetic ratio, 42.577 MHz per tesla: 600 / 42.577 ≈ 14.09 T. A 400 MHz magnet is about 9.40 T and a 1.2 GHz magnet about 28.2 T.

How do I measure a coupling constant J from peak positions in ppm?

Take the difference between two adjacent lines of a multiplet in ppm and multiply it by the spectrometer frequency in MHz. Lines at 7.2512 and 7.2302 ppm on a 400 MHz instrument are 0.0210 × 400 = 8.40 Hz apart. In a first-order multiplet equal spacings give J; unequal spacings mean several couplings or a second-order pattern.

How do I reference a spectrum to the residual solvent peak?

Find the residual signal of the deuterated solvent in your spectrum, compare it with the accepted value (for example 7.26 ppm for CHCl₃ in CDCl₃ ¹H, 77.16 ppm for CDCl₃ ¹³C), and add the difference to every peak. If CHCl₃ appears at 7.24 ppm, add +0.02 ppm to all shifts.