Nucleic Acid Concentration Calculator – From A260 to ng/µL, Copies and Purity
Enter the absorbance readings from a spectrophotometer and this nucleic acid concentration calculator returns the concentration of a DNA, RNA or oligonucleotide sample, the A260/A280 and A260/A230 purity ratios with a verdict, and the same amount as a molar concentration, copies per µL and pmol of ends. Everything is calculated in your browser.
Why nucleic acids absorb at 260 nm
The purine and pyrimidine bases absorb ultraviolet light with a maximum near 260 nm. The Beer–Lambert law links that absorbance to concentration: A = ε · c · l. For a mixed sample you rarely know ε, so a single conversion factor is used instead.
Where 50, 33 and 40 come from
One absorbance unit at 260 nm over a 1 cm path corresponds to about 50 µg/mL of double-stranded DNA, 33 µg/mL of single-stranded DNA and 40 µg/mL of RNA. The differences come from hypochromicity: stacked base pairs absorb less than free bases, so a duplex needs more mass than a single strand to give the same reading. These factors are averages over typical base composition, which is why the factor field is editable.
Oligonucleotides: why the sequence matters
A short oligo has too few bases for an average to be reliable, so the tool computes its ε260 from the sequence with the nearest-neighbour method. Each pair of adjacent bases contributes its own absorbance, and the single bases inside the chain are subtracted once. The concentration is then A260 × DF / (ε260 × l). A dye or other modification adds absorbance that the sequence cannot predict, so use the supplier's ε260 in the override field when you have it.
Reading 260/280 and 260/230
A 260/280 ratio near 1.8 is typical of pure DNA and near 2.0 of pure RNA. Lower values suggest protein or phenol. A 260/230 ratio of 2.0 to 2.4 is expected for a clean sample; lower values point to guanidine salts, EDTA, phenol, carbohydrate or TRIzol carry-over. The ratios depend on pH, reading lower in unbuffered water, so measure in a buffer such as TE. They are also unreliable at low concentration, below roughly 10 ng/µL of dsDNA.
Pathlength, dilution and the linear range
The tool corrects for the background reading, multiplies by the dilution factor and divides by the pathlength, so a 1:20 dilution read in a 1 cm cuvette is handled exactly. Most cuvette instruments are reliable for corrected A260 between about 0.1 and 1.5. Microvolume instruments switch their own pathlength and display a value normalised to 10 mm, so the linear-range check is skipped for them.
From ng/µL to copies
Copy number follows from the molecular weight: copies/µL = ng/µL × 10⁻⁹ × N_A / MW. For a double-stranded fragment of N base pairs the free-acid weight is about 617.9 × N g/mol. Many calculators use the sodium salt instead, about 661.9 g/mol per bp, which is where the rules of thumb of 650 and 660 come from. Choosing the sodium-salt basis changes the molar value and the copy number by about 7 percent, and never changes ng/µL.
When to use a fluorometric assay instead
A260 cannot tell DNA from RNA or from free nucleotides. For low-concentration samples, or when RNA contamination matters, a dye-based assay such as Qubit or PicoGreen is more specific; use a standard curve to convert its signal to a concentration.