Wheatstone Bridge Calculator

Physics

Balance mode

Inputs
R4 = R2 · R3 / R1

%

%

%

Only needed for node voltages, currents and power.
Result

R4 (Rx) at balance

1.034 kΩ

= 1034 Ω

Nearest E24 value

1 kΩ

−3.288 % from the exact value

Nearest E96 value

1.02 kΩ

−1.354 % from the exact value

Worst-case minimum

1.003 kΩ

Worst-case maximum

1.065 kΩ

RSS estimate (first-order)

± 17.91 Ω

± 1.732 %, assuming uncorrelated errors

Bridge sensitivity at balance, dVout/dR4 = Vs·R3 / (R3 + R4)²

0.002078 V/Ω

Sensitivity is highest when the ratio arms are equal (R1 = R2 and R3 = R4).

At balance with this supply

Both nodes sit at 6.875 V, so the meter reads zero whatever Vs is. Left arm current 31.25 mA, right arm current 6.649 mA.

ArmResistanceCurrentPower
R1100 Ω31.25 mA97.66 mW
R2220 Ω31.25 mA214.8 mW
R3470 Ω6.649 mA20.78 mW
R4 (Rx)1.034 kΩ6.649 mA45.71 mW
Circuit
+Vs = 10 VR1100 ΩR2220 ΩR3470 ΩR4 (Rx)= 1.034 kΩGA6.875 VB6.875 V

At balance V_A = V_B, so the galvanometer needle rests at the centre.

Output against R4

Vout = V_B − V_A as R4 sweeps from half to twice its balance value. The output crosses zero at the balance point and goes positive as R4 increases.

About This Tool

Wheatstone Bridge Calculator – Balance, Sensitivity and Strain Gauges

The Wheatstone bridge is four resistors arranged in a diamond, with a supply across one diagonal and a sensitive meter across the other. It was popularised by Charles Wheatstone in 1843 and is still everywhere: inside digital multimeters, in load cells and pressure sensors, in hot-wire anemometers and in the school-lab meter bridge. This Wheatstone bridge calculator covers all of those uses: it finds the unknown resistance at balance, analyses an unbalanced bridge with a real galvanometer, sizes quarter, half and full strain gauge bridges in mV/V, and works the slide-wire meter bridge experiment.

Two voltage dividers side by side

The easiest way to understand the bridge is as two voltage dividers fed by the same supply. On the left, R1 sits above R2 and their junction, node A, is at V_A = Vs · R2 / (R1 + R2). On the right, R3 sits above R4 and node B is at V_B = Vs · R4 / (R3 + R4). The meter reads the difference, Vout = V_B − V_A. With this sign convention the output goes positive when R4, usually the unknown Rx, gets larger.

The bridge is balanced when the two nodes sit at the same voltage. Setting R2 / (R1 + R2) = R4 / (R3 + R4) and cross-multiplying gives the famous condition R1 / R2 = R3 / R4, or equivalently R1 · R4 = R2 · R3. Solve it for whichever arm is unknown: with R1 = 100 Ω, R2 = 220 Ω and R3 = 470 Ω, the bridge balances at Rx = 220 × 470 / 100 = 1034 Ω.

Why null measurement is so accurate

Notice that Vs cancels out of the balance condition. If the supply drifts, both node voltages drift together and the meter still reads zero. A null measurement therefore depends only on the ratio arms, not on the battery and not on the meter's calibration: the meter only has to tell zero from not-zero. That is why bridges reach accuracies far better than a direct ohmmeter reading. The supply does matter for sensitivity, the output change per ohm near balance, dVout/dR4 = Vs · R3 / (R3 + R4)². It is largest when the arms on each side of the output are equal, which is why equal ratio arms are the textbook choice.

The calculator also propagates resistor tolerance. The worst case pushes every known resistor to its limit in the unfavourable direction; the root-sum-square figure is a first-order estimate that assumes independent errors. It also suggests the nearest E24 and E96 standard values for the result.

The unbalanced bridge and the galvanometer

Away from balance, the meter sees a source with a Thevenin equivalent of Vth = Vout and Rth = R1‖R2 + R3‖R4, because shorting the supply puts each divider pair in parallel. A galvanometer of resistance Rg therefore draws Ig = Vth / (Rth + Rg). The tool cross-checks this against a full nodal solution of the loaded circuit, which also gives the true node voltages and the current and power in every arm.

Strain gauge bridges: quarter, half and full

A strain gauge changes resistance by ΔR/R = GF · ε, where the gauge factor is about 2 for metal foil. At 1000 µε that is only 0.2 %, far too small to read with an ohmmeter but easy to see as a bridge imbalance. In a quarter bridge one arm is active and the output is Vout/Vs = x / (4 + 2x), which is slightly nonlinear: the error is about x/2, roughly 0.1 % at 1000 µε. A half bridge uses two gauges with opposite strain, such as the top and bottom of a beam in bending, and gives exactly x/2. A full bridge with four active gauges gives exactly x, four times the quarter bridge output, with automatic temperature compensation.

Reading mV/V
Load cells quote sensitivity in millivolts of output per volt of excitation. A 2 mV/V cell excited at 10 V gives 20 mV at full load, and a ratiometric ADC that measures against the same supply cancels supply drift.

The meter bridge experiment

The meter bridge replaces R1 and R2 with a uniform one-metre wire. A jockey slides along it until the galvanometer shows no deflection at length l; since resistance is proportional to length, X = R · l / (L − l). With R = 10 Ω and a balance at 40 cm, X = 6.667 Ω. Results are most reliable when the balance point lies near the middle of the wire, and swapping X and R for a second reading cancels much of the error from the end connections. If you know the end corrections α and β, enter them to get X = R · (l + α) / (L − l + β).

Keep currents small
Every bridge heats its resistors with I²R. In precision work and with strain gauges, self-heating shifts resistance and causes drift, so check the power per element and use the lowest excitation that gives a readable signal.

Frequently Asked Questions

Is the Wheatstone Bridge Calculator free?

Yes, Wheatstone Bridge Calculator is totally free :)

Can I use the Wheatstone Bridge Calculator offline?

Yes, you can install the webapp as PWA.

Is it safe to use Wheatstone Bridge Calculator?

Yes, any data related to Wheatstone Bridge Calculator only stored in your browser (if storage required). You can simply clear browser cache to clear all the stored data. We do not store any data on server.

How does this Wheatstone bridge calculator work?

It treats the bridge as two voltage dividers fed by the same supply: R1 over R2 sets node A and R3 over R4 sets node B, and the output is V_B − V_A. Balance mode solves R1·R4 = R2·R3 for the missing arm, unbalanced mode adds the Thevenin equivalent and a nodal solve when a meter is connected, strain mode applies the exact quarter, half and full bridge equations, and meter bridge mode uses the slide-wire ratio X = R·l / (L − l).

Why doesn't the balance condition depend on the supply voltage?

At balance both nodes sit at the same fraction of the supply, R2/(R1+R2) = R4/(R3+R4), so the supply cancels out of the comparison entirely. Doubling Vs doubles V_A and V_B together and the meter still reads zero. This is why null measurements are so accurate: a drifting or noisy supply only changes how sensitive the bridge is, not where the null is.

Why is a quarter bridge nonlinear when half and full bridges are not?

In a quarter bridge only R4 changes, so the current in the right arm changes with strain and the output is x/(4 + 2x) rather than exactly x/4. The error is about x/2, roughly 0.1 % at 1000 µε with a gauge factor of 2. In the half and full bridges the active arms change by equal and opposite amounts, the arm totals stay constant, and the output is exactly proportional to x.

What does mV/V mean on a load cell or strain gauge datasheet?

It is the bridge output per volt of excitation, Vout/Vs expressed in millivolts per volt. A 2 mV/V load cell excited at 10 V gives 20 mV at full scale. Quoting the ratio lets the same sensor be used with any excitation voltage, and ratiometric amplifiers that measure Vout against the same Vs cancel supply drift.

How should I choose the ratio arms for maximum sensitivity?

For a given unknown, the output change per ohm, Vs·R3/(R3 + R4)², is largest when the arms on each side of the output node are equal, so equal ratio arms (R1 = R2, R3 = R4) give the steepest null. The balance point is unchanged by this choice, but a sharper slope makes small imbalances easier to see on the galvanometer.

What are the main sources of error in a meter bridge experiment?

End resistance at the wire clamps and connecting strips, a wire whose resistance per centimetre is not perfectly uniform, contact resistance and parallax at the jockey, and heating of the wire by the current. Balancing near the middle of the wire, keeping the current brief, and averaging two readings with the gaps swapped all reduce these errors; known end corrections can be entered as α and β.